Scilog SciPres Monitor:
Available in five sizes, SciLog® SciPres® scalable pressure sensors are designed for incorporation into single-use purification platforms. Pre-calibrated and gamma stable, calibration and identity information is stored on the sensor’s chip. The SciLog® SciPres combines pressure sensing capabilities and the convenience of disposability with easy set up. Each sensor is pre-programmed with a unique ID for easy traceability and data documentation when combined with the SciLog® SciDoc software. Factory calibration data is also stored on each sensor’s chip for out-of-box plug and play use.
PendoTech PressureMAT:
PendoTECH’s Single Use Pressure Sensors™ measure static and dynamic pressure of gases and liquids in your biopharmaceutical processes – accurately and cost effectively. Robust enough to be repeatedly cleaned and inexpensive enough to be utilized in disposable applications, PendoTech’s Single Use Pressure Sensors deliver a dependable alternative to stainless steel pressure transducers. The sensors feature PendoTECH High Accuracy Pressure (MEMS-HAP™) Chips and are perfect for filtration and chromatography processes, monitoring of glass and disposable bioreactors, filling operations and more. They are available in caustic-resistant polysulfone to withstand sanitization processes. PendoTECH’s pressure sensors, which are qualified for use up to 75 psi, are compatible with PendoTECH’s PressureMAT™ monitor/transmitter, PendoTECH’s Process Control Systems or other pre-qualified third-party monitors.
Tech Specs
SCILOG SCIPRES | SINGLE-USE BIOPHARM PRESSURE SENSORS |
---|---|
Operating Pressure | -5 to 60 psi |
Materials of Construction | Medical grade polysulfone |
Specifications Met | USP Class VI, FDA 21CRF177.1520 |
Connection Type | Luer, 3/8" Barb, 1/2" Barb, 3/4 Tri-Clover or 1" Tri-Clover 'Ladish' |
PENDOTECH PRESSUREMAT | PRESSURE SENSORS |
Pressure Range | -11.5 to 75 psi (-0.79 to 5.2bar) |
Biocompatability | All materials in contact with product fluid path meet USP Class VI requirements, both pre and post gamma irradiation |
Manufacturing Environment | FDA Registered, ISO 13485 certified facility; Class 7 clean room |
Accuracy | 0 to 6 psi: ±2% of reading. 6 to 30 psi: ±3% of reading. 30 to 60 psi: Typically better than ±5% of reading. 0 to -7 psi: ±3% of reading. -7 to -10 psi: ±5% of reading. |
Connector | Custom molded 4 pin connector; signal +/- and excitation +/ |